BIOBASE biochemistry analyzer is fully automatic,discrete,random access biochemistry analyzer.It is eupipped with:60 sample position,90 reagent position,and 120 reaction cuvettes.It is 400 test per hour,with both CE & FDA certified.
1.Sample unit:
Sample Settings: 60 sample positions, including calibration, quality control and emergency,
primary tubes and serum is available.
Sample Volume: 2ul ~50ul (0.1ul stepping).
2.Reagent unit:
Reagent Position: R1 45 & R2 45,total 90 refrigerated reagent position.
Reagent Volume: R1:25~300ul, R2:10~150ul (1ul stepping), supporting reagent of different brand
Temperature Control: continuous cooling system working within 24 hours.
- Reaction unit:
Cuvettes: 120 independent reaction cuvettes (optical path length 6mm)
Cleaning: automatic water cleaning and drying system.
Temperature Control: 37C±0.1C
4.Calibration and QC:
Have reaction curve and curve graph to check.
Calibration: one-point, multi-point
Linear Range: 0.000~2.500Abs
Quality Control: inserting quality control at random.
Parameter:
OverallPerformance | Model | BIOBASE-DIAMOND | ||||
Type | discrete, fully automatic | |||||
Throughput | 400 T/H | |||||
Analysis Method | End-point, Fixed-time,Rate | |||||
Programming Items | programmable for any test items | |||||
Certification | CE, FDA, ISO9001, ISO14001, ISO13485 | |||||
Sample unit | Sample Settings | 90 sample positions, including calibration, quality control and emergency. | ||||
Sample Volume | 2ul ~50ul (0.1ul stepping) | |||||
Sample Probe | Having liquid level induction function. | |||||
Probe cleaning | Automatic washing both interior and exterior | |||||
Analysis Items | analyzing maximally 45 items at the same time | |||||
Reagent unit | Reagent Position | R1:45 R2:45 total 90 refrigerated reagent position | ||||
Reagent Volume | R1:25~300ul, R2:10~150ul (1ul stepping), supporting reagent from different brand | |||||
Temperature Control | continuous 2C~8Ccooling within 24 hours | |||||
Reagent Probe | Having liquid level induction function | |||||
Probe cleaning | Automatic washing both interior and exterior | |||||
Reaction unit | Reaction Time | 10 minutes | Temperature | 37C±0.1C | ||
Cuvettes | 120 reaction cuvettes (optical path length 6mm) | |||||
Mixer | after injecting reagent and sample, blending immediately | |||||
Cleaning | automatic water cleaning and drying system | |||||
System Adjustment | System record the blank value and get more accurate result by deducting blank base value of each cuvette | |||||
Optical System | Light Source | 12V/30W halogen-tungsten lamp, 2000 hours | ||||
Receiver | 8 precision photo receivers | |||||
Wavelength Range | 340,405,450,510,546,578,630,700mm | |||||
Calibration and QC | Calibration | one-point,multi-point | Linear Range | 0.000~2.500Abs | ||
Quality Control | inserting quality control at random. | |||||
Software | Emergency Function | emergency sample can be arranged at any time | ||||
Software | Windows XP or 7, sample and easy to operate | |||||
Data Processing | Having reaction curve detection and data display function in whole process. | |||||
Others | Power Supply | AC220V+22V/50Hz+1Hz | ||||
Dimension | 1070mm×910mm×1230mm (W*D*H) | |||||
Weight | 180kg |