Model |
AC9802 |
|||
Measuring item |
K+、Na+、Cl-、Ca、pH |
|||
Calculating item |
nCa、TCa |
|||
Measuring parameter |
Measuring |
Measuring scope |
Resolving power |
Repeatability |
item |
(mmol/L) |
(mmol/L) |
error |
|
(K+) |
0.50~15.00 |
0.01 |
CV≤1.0% |
|
(Na+) |
30.0~200.0 |
0.1 |
CV≤1.0% |
|
(Cl-) |
30.0~200.0 |
0.1 |
CV≤1.0% |
|
(Ca2+) |
0.10~5.00 |
0.01 |
CV≤1.0% |
|
pH |
4.0~9.0 |
0.01 |
CV≤1.0% |
|
Measuring time |
Measuring time: 25s |
|||
Time for sampling, measuring, washing and printing: 40s |
||||
Sample size |
100μL(manual positioning 60μL) |
|||
Data storage |
10000 |
|||
It will refresh automatically if it is full |
||||
Communication |
232 interface |
|||
interface |
||||
Display |
240×64 LCD |
|||
Printer |
58mm thermal printer |
|||
Power |
AC220V±22V 50Hz 25W |
|||
supply |
||||
Size |
408mm×390mm×220mm |
|||
Weight |
5.5Kg |
|||
Measuring |
ISE |
|||
method |
||||
Measuring |
Temperature: 10℃~32℃ |
|||
condition |
Relative humidity ≤85% |