Project Description
Principle of Luminescence
•Non-enzyme involved flash chemiluminescence, long stability of reagents
•ABEI label, stable in acid and alkaline solution
Main Features
•Throughput: up to 180 tests/hour
•24 hours ready to use
•Time to first result:17 minutes
Modes of Operation
•Random, Batch and STAT
Sample Loading
• Up to 40 sample tubes
• Continuous loading, STAT available
• Analyzer auto numbered
• LIS connection
• Refrigerated sample area with independent power supply
Reagent Loading
•9 reagents on board
•Continuous loading
•RFID reading all info of reagents
•Refrigerated reagent area
Reagent Features
•Integrated kit, ready-to-use, no pretreatment required
•Integrated kit including calibrators
•Using superparamagnetic microbeads
•RFID tag storing all info of reagent
•RFID tag with built-in master curve
•2-point calibration to adjust master curve
•Calibration stability: Max 4 weeks
Other Features
•Clot detection
•Liquid level detection for reagent
•Optional ratio for sample dilution
•Auto dilution for high concentration sample
•Constant 37℃ incubation
•15-30℃ Operating Temperature
Operation System
•Windows 7
•HP compatible (Dual-core CPU), Color touch screen
Interconnection
•Connection to LIS via COM or network card
•Direct operation on LIS system to test
Dimensions Weight
•102x72x56cm, 73Kg